4-20mA Transmitter for dissolved oxygen sensors, DIN rail

€380.00
Net: €319.33

Prices incl. VAT plus shipping costs

On stock.

Ask for Offer
  • 35402
UMS Transmitter, DIN rail Measuring amplifier for automation. compatible with all... more
Product information "4-20mA Transmitter for dissolved oxygen sensors, DIN rail"

UMS Transmitter, DIN rail

Measuring amplifier for automation.

  • compatible with all UMS DO sensors and DO sensors for ink
  • Measuring ranges : oxygen in mg / l and %sat ( switchable )
  • Fully automatic temperature compensation
  • Automatic pressure compensation with integrated air pressure sensor for higher accuracy during calibration and to calculate the relative value (% saturation)
  • Only 1 unit wide on the DIN rail
  • 4 - 20 mA current loop
  • 24 mA failure signal
  • Power supply via current loop ( 2 wire / loop powered ) or separately ( 4 wire )
  • Easy Calibration by pushbutton using the calibration chamber
  • Synchronous sampling according to the mains frequency (50 and 60 Hz) increases immunity to interference. (Especially while using long sensor cable and parallel routing to power cables)
  • Fault detection temperature sensor (short circuit or not connected)
  • also available in waterproof case
  • extended measuring range available on request

Two display LEDs:
1. Status LED :  permanently indicates the selected measuring parameter. (% Sat or mg / l)
2. Error LED : indicates type of an error occurring

Operating manuals, software, drawings, sample programs and product flyers are available in the download area.

 

Measurment ranges Accuracy
0.0 .. 19.9 mg/l 0.02 mg/l
0.1 .. 199.9 %sat 0.1 %sat

 

Output current 4 - 20 mA
Failure signal 24 mA
Sampling rate 1 Hz
Supply voltage 9 - 28 V DC
Dimensions ( LxBxH ) 90 x 17.9 x 59



Related links to "4-20mA Transmitter for dissolved oxygen sensors, DIN rail"
Viewed